https://www.selleckchem.com/products/gpna.html To investigate the suitability of the microDiamond detector (mDD) type 60019 (PTW-Freiburg, Germany) to measure the anisotropy function F(r,θ) of High Dose Rate (HDR) Ir brachytherapy sources. The HDR Ir brachytherapy source, model mHDR-v2r (Elekta AB, Sweden), was placed inside a water tank within a 4F plastic needle. Four mDDs (mDD1, mDD2, mDD3, and mDD4) were investigated. Each mDD was placed laterally with respect to the source, and measurements were performed at radial distances r=1cm, 3 and 5cm, and polar angles θ from 0° to 168°. The Monte Carlo (MC) system EGSnrc was used to simulate the measurements and to calculate phantom effect, energy dependence and volume-averaging correction factors. F(r,θ) was determined according to TG-43 formalism from the detector reading corrected with the MC-based factors and compared to the consensus anisotropy function F(r,θ). At 1cm, the differences between measurements and MC simulations ranged from -0.8% to +0.8% for θ=0° and from -2.1% to+2.3% for θ≠0ating F(r,θ) of an HDR 192 Ir brachytherapy source in a water phantom using the mDD. The phantom effect and the volume-averaging need to be taken into account, especially for the smaller distances and angles. Good agreement to CON F(r,θ) was obtained. The discrepancies at (1 cm, 0°), accurately predicted by the MC results, may suggest a reconsideration of CON F(r,θ), at least for this position. The slight overestimations at (3 cm,0°) and (5 cm,0°), both in comparison to CON F(r,θ) and MC results, may be due to an underestimation of the air volume between source and needle tip, dark current, intrinsic over-response of the mDDs, or radiation-induced charge imbalance in the detector's components. The results indicate that the mDD is a valuable tool for measurements with HDR 192 Ir brachytherapy sources and support its employment for the determination and validation of TG-43 parameters of such sources.The present research aims to enha